Precautions:
It is forbidden to spray alcohol directly to the air outlet at the rear of the machine cavity and the two ends of the ultraviolet lamp, which will directly cause damage to the cross-flow fan and the ultraviolet lamp.
Do not look directly at the UV lamp for a long time.
The box is equipped with UV lamp with sterilization function. Pay attention to the ultraviolet lamp radiation when using it, and it must be turned off when not in use.
The equipment should be moved in parallel, and the maximum tilt of the box cannot exceed 45 degrees.
Installation requirements: The incubator should be placed on a solid, flat surface; strongly corrosive gas and relative humidity of about 85%; good indoor ventilation;
the distance between the back of the oscillator and the wall should be greater than 20 cm; the right side needs to be cooled within 40CM, and there must be no obstructions;
When the equipment works in the "cooling" state for a long time or when it is stopped for more than three months, it should be heated as scheduled.
The tide-repellent treatment is performed every two weeks. The method is to reset the temperature to 40 ° C and open the cover to make the equipment work in the "heating" state. Each time cannot be less than 4 hours, so that the moisture in the circulation box is fully dispersed.
Do not place flammable or explosive materials.
Keep the inside of the box clean, and often clean up debris.